RS-10D Calibration Light Source

The RS-10D calibration light source is a precision source of radiant flux, used primarily to calibrate light-measuring instrumentation and as stimuli to measure detection devices.

Conservative design and rigid quality standards make Gamma Scientific sources the ones by which others are judged. They may be used as standards of irradiance or radiance, traceable to the National Institute of Standards Technology (NIST).

Near-constant radiant flux output is achieved via exclusive use of tungsten-halogen lamps. Ultra-stable, constant-current supplies utilize precision shunt-current measurement and comparison circuits built into the source housing.

With RS-70-X accessories, the RS-10D light source provides an absolute reference for producing standard outputs in a complete range of luminance, illuminance, spectral-radiance and spectral-irradiance units. Additional special filters and attenuators enable generation of different color temperatures, spectral outputs and intensities. Calibrations in photometric and/or spectroradiometric units are supplied and are traceable to NIST.

Request a Quote
Features
  • NIST traceable
  • Built-in timer shows elapsed time
  • 200-hour calibration/one year
  • Calibrate spectroradiometers, radiometers and photometers
  • Calibrate detector responsivity
  • Measure reflectance and transmittance
  • Tungsten-halogen lamps for stable output
  • Calibration reports in units of luminance, illuminance, radiance and irradiance

Download Data Sheet (PDF)

Download Now

RS-10D Spectral Irradiance Head

Calibrated Wavelength Range @ Data Interval300-1100 nm @ 5 nm
Calibration Interval200 hr or 1 year
Output, Nominal Irradiance at 25 cmW/cm2*nm
8.0 x 10-9 at 300 nm
7.5 x 10-7 at 550 nm
2.0 x 10-6 at 800 nm
2.2 x 10-6 at 1100 nm
Output, Nominal Illuminance at 25 cm592 lm/m2 (LUX); 55 lm/ft2 (fc)
Nominal Radiance with RS-70-2 DiffuserW/cm2*sr*nm
6.2 x 10-9 at 300 nm
1.0 x 10-6 at 550 nm
9.0 x 10-6 at 800 nm
6.5 x 10-5 at 1100 nm
Nominal Luminance with RS-70-2 Diffuser850 cd/m2
Correlated Color Temp. (CCT)2856 ± 20K Reported to ± 7K
Correlated Color Temp. with RS-70-2 Diffuser2575 ± 50K Reported to ± 7K
Uniformity of Diffuser±3% over 25 mm
NIST Standards of Irradiance Output Uncertainty±2.5%
Uniformity of Exit Port±0.75%
Uniformity of Irradiance Plane Normal to Exit Port±2% over 50 mm; diameter at 25cm
SizeHeight: 203 mm (8.0 in)
Length: 197 mm (7.8 in)
Width: 175 mm (6.9 in)
Weight: 2.8 kg (6.25 lbs)
Regulator TypeConstant Current
Output Current4A Maximum
Current Accuracy, Long TermBetter than 0.5%
SettabilityBetter than 0.2%
Temperature DriftLess than ±.25%/10°C
Temperature Range15°C to 35°C
Humidity10%-85% non-condensing
RegulationLess than ±0.2% change for 10 Volt line change
Thermal drift After 8 Minute WarmupLess than 0.1%
Current Ramp On/Off TimeApproximately 30 seconds
Power30 Watts Maximum
Line Voltage100-240VAC, 50-60 Hz