GS-191-FA-1045 Rotary Reflectance Measurement System

The GS-191-FA-1045 is a rotary-loaded gonio-reflectance measurement system that automatically captures complete spectral and colorimetric properties for coated glass, polished substrates or diffuse surfaces at 10° and 45° angle of incidence simultaneously, without requiring secondsurfacemasking. Depending on system model, substrates as thin as 0.35 mm can be tested.

Based on high precision spectroradiometric instrumentation, proprietary measurement techniques and expertise in low-light measurement technology developed by Gamma Scientific, the product range features industryleading accuracy, repeatability and throughput, including both refractive index determination and thin film coating thickness.

Applications

  • Anti-reflectance coating characterization
  • Flat-panel display glass testing
  • Touchscreen display glass testing
  • Optical filter / lens testing
  • Pyrolytic glass coating test & characterization
  • Characterization of flat panel displays, photovoltaic coatings, low-E architectural coatings, paint samples and diffuse plastics
Request a Quote

Highly Accurate & Repeatable Reflection Measurements

  • Nondestructively capture complete spectral and colorimetric properties with scan times as short as 200 msec per measurement point
  • Isolated first-surface measurement of thin glass substrates down to 500 µm in thickness without requiring second-surface masking
  • Measure total reflectance or isolate internal optical interfaces
  • Test capability for diffuse or specular surfaces
  • Programmable, multi-location measurement, pass/fail criteria settings and binning capabilities
  • Configuration options including handheld, semi-automatic and fully automatic rotary systems with robotic loading

In addition to our exceptional technical and functional capabilities, Gamma Scientific is ISO/IEC 17025 accredited by NVLAP (NVLAP lab code 200823-0).

Power Supply

Voltage 230V ±5%
Frequency 50/60 Hz ±3Hz
Maximum Ouput Power 1000 W
Rated Power 500 W
Rated Current 2.5 A

Vacuum

Maximum Pressure -70 (kPa)
Flow Capacity >80 (L/min) @-80 (kPa)

Weight

1000 kg (Rotary) 1m open door space reserved around

Environment

Ambient Temperature 10 to 35° C
Relative Humidity <90% non-condensing
Dust <0.2 mg3
Vibration <0.98 m/s2 (0.1G), 10-60 Hz

Optical Specifications

191 Optical Head

(Optional Geometries of 0° and 20°)

10 Degree Angle of Incidence

45 Degree Angle of Incidence

Measurement Type

First surface specular reflection

First surface specular reflection

Sample Types

Glass

Glass

Illumination Angle

 10°

  45°

Viewing Angle

10°

45°

Maximum Sample Thickness

(first-surface reflectance only)

0.5 mm or thicker

0.35 mm or thicker (GS-191-FA-1045-V)

0.25 mm (transparent samples)

Maximum Sample Thickness

6 mm

6 mm

Maximum Sample Size

280 mm (11") diagonal

Spectral Range

360 to 830 nm

360 to 830 nm

Illumination Spot Size (sample area)

1 mm x 10 µm

1 mm x 10 µm

Measurement Speed (typical)

< 1500 msec

< 1500 msec

Calibration Reference Standard

Integral BK-7 polished glass

Integral BK-7 polished glass

Spectral Reflectance

± 0.5%

± 0.5%

Tristimulus (CIE 1931 X,Y,Z)

± 0.05

± 0.10

Chromaticity (CIE 1931 x,y)

± 0.005

± 0.005

LAB Color (CIE 1976 L*, a*, b*)

L ± 2.0                       a*, b* ± 0.8

L ± 2.0                       a*, b* ± 0.8

Average Reflectance

± 0.2

± 0.2

 

Rotary System Specifications

191 Optical Head

191F-1045 Dual Angle Optics

Measurement Locations per Panel

3 measurement locations standard. Configurable up to 5 locations. Configurable for 10 and 45° measurement at any location.

Glass stations

8 glass stations; 3 for loading / unloading and 5 for measurement optics.

Measurement Cycle Time

< 7 seconds per panel 

Sample Load/Unload Cycle Time

< 3 seconds per panel, operator dependent

Spectral Data

Reflectance as a function of wavelength

Colorimetric Data

Tristimulus 1931 X,Y,Z                   Tristimulus 1964 X,Y,Z                    CIE 1931 x,y         

CIE 1976 L*, a*, b*                         CIE 1976 L*, u*, v*

System Dimensions

1.75 meters H x 1.6 meters W x 1.6 meters D     Weight 1000 kg

Operating Ranges

Ambient Temperature  0 to 35°C                           Relative Humidity < 90% non-condensing