$1,900.00
The MK350N Premium is a lightweight, portable and easy-to-use spectral illuminance (lux) meter or spectral light meter. It is designed for fast and accurate illuminance measurements for general lighting and LED applications. Designed to be intuitive and user-friendly, it is ideal for measuring light parameters such as lux, CCT, CRI, CIE Chromaticity Coordinates, Ra, R1 to R15, wavelength and others. The device also includes an integrated flicker meter.
Precision, Power and Portability in a Compact & Lightweight Package
- Completely Stand-Alone Operation - no other equipment necessary (e.g., PC, Smartphone, etc.)
- More than 40 units of measurement
- Built-in file browser allows for quick access to previously saved data
- 3.5” color touch screen with intuitive menu selection
- Conforms to ISO 14001, JIS, DIN and IECQ standards
- Wi-Fi remote control option with IOS or Android applications
- Automatic continuous measurements with data-save to an SD card
- NIST Traceable calibration
Product Features |
||
Capture Function |
One-time or continuous |
|
Operation Mode |
Stand-alone, WiFi or USB (to PC, mobile device or tablet) |
|
Integration Mode |
Auto or Manual |
|
Automatic Dark Calibration |
Auto mode |
|
Measuring Modes |
Basic Spectrum CRI CIE 1931 Chromaticity CIE 1976 Chromaticity TM30-15 Flicker Frequency Browser Option |
|
Measuring Capabilities |
Light & Color Parameter |
Illuminance (lux) or Foot-candle (FC) Correlated Color Temperature (K) CIE Chromaticity Coordinates CIE 1931 x,y CIE 1976 u’, v’ CIE 1931 XYZ ΔX, ΔY, ΔU’, ΔV’ Delta UV; DUV Dominant wavelength; Hue, ƛd (nm) Chroma Purity (%) Scotopic and Photopic ratio: S/P |
Light & Color Evaluation |
Color Rendering Index (CRI, Ra) R1 to R15 Color Quality Scale (CQS) Gamut Area Index (GAI) TM-30-15 (Rf, Rg, Color Vector Graphic) Television Lighting Consistency Index (TLCI) |
|
Flicker |
Flicker Frequency (Hz) Flicker Percentage (%) Flicker Index Stroboscopic Effect Visibility Measure (SVM) |
|
Spectral Radiation |
Spectral Power Distribution (SPD) mW/m2 Peak Wavelength (ƛp) nm Peak Wavelength Value (ƛpV) mW/m2 Integration time (I-time) Scotopic and Photopic ratio (S/P) |
Spectrum |
||
Sensor |
CMOS linear image sensor |
|
Wavelength Range |
380 to 780 nm |
|
Wavelength Data Increment |
1 nm |
|
Spectral Bandwidth |
12 nm (half power bandwidth) |
|
Wavelength Reproducibility |
± 1 nm (assumes stable input light source) |
|
Measurement Range |
5 to 100,000 lux |
|
Illuminance (1)
[From 0.05 to 5,000 cd/m2] |
Accuracy |
± 5% |
Repeatability (2σ) |
± 0.2% from 100 to 100,000 lux ± 0.5% from 5 to 100 lux |
|
Color (1,2)
|
Accuracy |
x y: ± 0.002 from 100 to 100,000 lux x y: ± 0.0025 from 5 to 100 lux |
Repeatability |
± 0.0005 in CIE 1931 x,y |
|
Repeatability (2σ) |
x y: ± 0.0002 from 500 to 100,000 lux x y: ± 0.0004 from 30 to 500 lux x y: ± 0.001 from 5 to 30 lux |
|
CCT Accuracy |
± 2% |
|
CRI Accuracy at Ra |
± 1.5% |
|
Stray Light |
-25 dB maximum (550 ± 40nm monochromatic source) |
|
Integration Time Range |
100 µsec to 1 sec |
|
Digital Resolution |
16 bit |
|
|
||
Flicker |
||
Measurement Range |
5 to 100,000 lux |
|
Sampling Rate |
100 kHz |
|
Frequency Range |
5 to 50 kHz |
|
Frequency Resolution |
2, 4, 8, 16, 32 Hz |
|
Flicker Accuracy |
± 5% |
|
|
||
System Configuration |
||
Display |
320 x 240 (3.5 in) resistive touch LCD |
|
Maximum Files |
68,000 with 8 GB SD card, compatible with Excel® and JPG |
|
Battery Operation |
Up to 5 hours, onboard 3.7 V Li-ion |
|
External Power |
Adapter (included), 2500mAh via USB connector |
|
Data Interface |
SD card (SD2.0.SDHC up to 32 GB) or mini USB port (USB 2.0) or WiFi SD card (IOS or Android) |
|
Dimensions |
148 mm (5.9 in) H x 78 mm (3.1 in) W x 24 mm (.95 in) D 225 g (0.5 lbs) including battery |
|
Language Options |
English, Traditional Chinese, Simplified Chinese, Japanese, Spanish, German, French, Italian, Russian |
|
Camera Resolution |
2M pixels |
|
(1) At 23 ± 2° C and relative humidity ≤ 50% Specifications are subject to change without notice. (2 )Illuminant A at 2,856 K at 20,000 lux (3) 30 Hz AC/DC 10% sine wave unless otherwise specified |