For the ultimate in resolution and accuracy, the SpectralLED® 940-SWIR source incorporates 16 discrete LED devices, all with a center wavelength of 940 nm. The output is directly coupled into a 38mm diameter homogenizing rod, with a length of 190 mm, having a diffuser optically bonded to the output surface.
The platform is easily adaptable for automated test systems and production line integration, with integrated optical feedback and temperature control to ensure rock-solid stability and consistent results.
Applications
- Camera and image sensor calibration
- Facial recognition system test & calibration
- Spectrum / illuminant simulation
- Diagnostic imaging
Unprecedented Resolution and Accuracy for Facial Recognition System Test & Calibration
- Constant current drivers & built-in optical feedback
- Accurate & flicker-free output in real-time
- All solid-state design for rapid start-up, repeatable performance
- ISO/IEC 17025 accredited by NVLAP (NVLAP lab code 200823-0) for calibration accuracy
Please contact us with your custom & OEM requirements.
RS-7-940 Optical Specifications |
|
---|---|
Spectral Range |
940 ± 10 nm (Custom ranges available on request) |
Spectral Output |
16 discrete LED channels |
Spectral Bandwidth |
Typical 50nm FWHM |
Source Geometry |
38mm diameter homogenizing rod, with a length of 190 mm, having a diffuser optically bonded to the output surface. |
Spatial Uniformity |
≥ 97% |
Calibration |
Irradiance (µW / cm2) at the plane of the diffuser output surface |
Accuracy Specifications |
|
Illumination Stability |
≥ 99.99% after 50 ms |
Illumination Accuracy |
± 1% Absolute, NIST traceable |
Spectral Accuracy |
± 1 nm centroid wavelength |
Linearity |
< 0.1 % RMS of full scale |
Temperature Stability |
Within ± 1° C via active TEC |
Long-term Drift |
Output ≤ 2% Spectral ≤ 1 nm (typical, channel dependent) |
Electrical Specifications |
|
Electrical Resolution |
16 bit DAC for channel current drivers 24 bit ADC for internal radiance monitor feedback |
Dynamic Range Adjustment |
4-5 decades typical (spectrum dependent) |
LED Control |
Pure DC constant current with floating differential sensing |
General Specifications |
|
Software |
Firmware includes full spectral calibration with spectral fitting, preset storage, real-time optical feedback, radiometric units supported |
Interface Connectors |
USB 2.0 type B and DB-9 |
Interface Protocol |
Simple ASCII commands with optional binary block transfer |
Supported Operating Systems |
USB drivers for Windows, OSX and Linux via FTDI virtual COM port Legacy RS-232 serial port for integration (no OS required) |
Input Voltage & Power |
100 to 240 VAC at 50-60Hz, 600W maximum |
Dimensions (H x W x L) |
405mm (16 in) x 460mm (18.1in) x (305mm (12in). Weight 17.5 kg (38.6 lbs) |
Options |
|
RS-7 Wavemon™ Upgrade |
Multi-channel photodiode system w / amplitude feedback & real-time ƛ measurements |
RS-7 Wavemon™ Multi-channel photodiode system with amplitude feedback and real-time wavelength measurements